{"url_path":"/fda/device/recalls/Z-3079-2026","section_key":"summary","section_title":"DEVICE Recall: Medline Industries, LP — Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators    ORTHO ASPIRATION KIT\t  \tDYKM1636A","topic":"fda","document":{"doc_type":"device-recall","doc_date":"2026-07-20","source_url":"https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3079-2026%22","accession_number":null,"cik":null,"ticker":null,"issuer_name":null,"edgar_url":null,"primary_entity_key":null,"primary_entity_name":"Medline Industries, LP"},"word_count":103,"has_tables":false,"body_markdown":"# FDA DEVICE Recall: Z-3079-2026\n\n**Recalling firm:** Medline Industries, LP\n**Classification:** Class II\n**Status:** Ongoing\n**Type:** Voluntary: Firm initiated\n**Initiated:** 2026-07-20\n**Report date:** 2026-09-09\n\n## Product\nMedline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators    ORTHO ASPIRATION KIT\t  \tDYKM1636A\n\n## Reason\nDownstream recall - manufacturer recalled specific lots of BD ChloraPrep Clear - 1mL Applicators. Affected product may exhibit an open or incomplete seal on the packaging of the applicator. Use of potentially contaminated applicator may lead to localized skin or soft tissue infection, or systemic infection or infectious complications.\n\n## Affected lots\nUDI-DI 10889942925858   Lot 26EBI940\n\n## Distribution\nUS Nationwide distribution."}