# MULTISCAN PHOTON ABSORPTIOMETRY & OSTEOPOROTIC FRACTURE

> **NIH NIH R01** · BETH ISRAEL DEACONESS MEDICAL CENTER · 1991 · $148,364

## Abstract

Approximately 1.3 million osteoporotic fractures of the vertebrae hip,
and distal radius occur each year in the U.S., with estimated annual
costs in excess of $10 billion.  Despite this, the biomechanics of the
osteoporotic state have been little studied and controversy exists on the
most suitable procedure to identify individual patients at high fracture
risk.  Until relatively recently, the most widely accepted screening
procedure for osteoporosis was Single Photon Absorptiometry (SPA) to
measure bone mineral content (BMC) of the distal radius.  However, a
number of studies have shown BMC of the distal radius to be a relatively
weak predictor of fracture of the hip and spine and thus, in recent
years, dual photon absorptiometry (DPA) and dual energy x-ray
absorptiometry (DEXA) of the spine, hip, and total body and QCT of the
spine have replaced SPA in many centers. Despite these trends, single
photon absorptiometry remains an attractive option for clinical
osteoporosis screening, primarily because of tits modest cost, low
radiation dose, long-term precision, and large installed instrument base.
In recent studies, we have identified biomechanical factors that we
believe help explain the relatively poor performance of SPA of the distal
radius to discriminate between fracture and non-fracture groups at remote
skeletal sites.  supported by an NIH Biomedical Resource Technology
Program for Pilot Studies, we have developed a new Multiscan Single
Photon Absorpitometry (MSPA) procedure for the distal radius which allows
extraction of the necessary biomechanical information by performing
multiple scans at varying angular orientations.  Pilot experiments with
cadaveric forearms have demonstrated highly significant improvements
(when compared to conventional SPA) in the prediction of in-vitro failure
loads necessary to simulate a Colles' fracture in-vitro.  Our first
objective in the proposed investigation will be to extend these pilot
experiments to a sufficiently large sample to establish the precision and
accuracy of the technique.  Our second objective will then be to relate
the MSPA measures and failure loads at the distal forearm to DEXA
parameters and fracture loads for the proximal femora from the same
cadaveric specimens.  These correlations will test the hypothesis that
radial MSPA parameters can be used to predict fracture risk at the
proximal femur under controlled in-vitro loading conditions.  And
finally, to begin to explore the potential clinical applications of
Multiscan Single Photon Absorptimetry, we will conduct a prospective
comparison between MSPA parameters obtained at the distal radius and DEXA
parameters obtained at the hip in patients being screened for
osteoporosis.  Through this comparison of non-invasive procedures applied
at the wrist and hip, using a combination of in-vitro testing and
clinical evaluation, we hope to provide a simple and inexpensive
procedure which can be implemented on commercially available forearm
scanners...

## Key facts

- **NIH application ID:** 3159992
- **Project number:** 5R01AR039760-03
- **Recipient organization:** BETH ISRAEL DEACONESS MEDICAL CENTER
- **SAM.gov UEI:** C1CPANL3EWK4
- **Principal Investigator:** WILSON C HAYES
- **Activity code:** R01 (R01, R21, SBIR, etc.)
- **Funding institute:** NIH
- **Fiscal year:** 1991
- **Award amount:** $148,364
- **Award type:** 5
- **Project period:** 1989-07-20 → 1993-11-30

## Primary source

NIH RePORTER: https://reporter.nih.gov/project-details/3159992

## Citation

> US National Institutes of Health, RePORTER application 3159992, MULTISCAN PHOTON ABSORPTIOMETRY & OSTEOPOROTIC FRACTURE (5R01AR039760-03). Retrieved via MORVS 2026-10-24 from https://api.morvs.ai/grant/nih/3159992. Licensed CC0.

---

*[NIH grants dataset](/datasets/nih-grants) · CC0 1.0*