FDA recall Z-0119-2020

Horiba Instruments Incorporated · Class II · device

Product

Clinical Chemistry, ABX Pentra Lipase CP, ABX Pentra 400/Pentrac400, REF: A11A01631, Model/Part # 1220001631, UDI # 03610230004944

Reason for recall

Potential for chemical analyzer to report abnormal results due to an interference with triglycerides CP and Cholesterol CP. An over estimation of the lipase concentration could be observed and a patient result could be declared abnormal by the instrument due to these interferences instead of normal result < 38 U/L (or < reference value used in your laboratory).

Distribution

US: AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY

Key facts

Status
Terminated
Initiation date
2018-10-30
Report date
2019-10-16
Termination date
2020-06-16
Voluntary/Mandated
Voluntary: Firm initiated
Location
Irvine, CA, United States