FDA recall Z-0664-2025

Micro-X Ltd. · Class II · device

Product

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Reason for recall

Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.

Distribution

Worldwide and US Nationwide Distribution: NJ, PR, CA

Key facts

Status
Ongoing
Initiation date
2023-05-31
Report date
2024-12-18
Voluntary/Mandated
Voluntary: Firm initiated
Location
Tonsley, Australia

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0664-2025