FDA recall Z-0848-2022

RAYSEARCH LABORATORIES AB · Class II · device

Product

TomoHelical and TomoDirect plans in RayStation 6, 7, 8A, 8B, 9A, 9B, 10A, 10B and 11A. Build Numbers: 6.1.1.2, 6.2.0.7, 6.3.0.6, 7.0.0.19, 8.0.0.61, 8.0.1.10, 8.1.0.47, 8.1.1.8, 8.1.2.5, 9.0.0.113, 9.1.0.933, 9.2.0.483, 10.0.0.1154, 10.0.1.52, 10.1.0.613, 11.0.0.951 or 11.0.1.29.

Reason for recall

If the dynamic jaw mode is used and an optimization is continued after changing the dose grid or modifying the target ROI, the positions of the jaws may change unexpectedly for some control points. User must be aware to avoid unnecessary jaw openings during treatment planning

Distribution

US Nationwide distribution in the states of AL, CA, FL, GA, IA, IL, MO, MS, NC, OK, OR, SC, TN, WA, WI.

Key facts

Status
Terminated
Initiation date
2021-09-09
Report date
2022-04-06
Termination date
2024-04-16
Voluntary/Mandated
Voluntary: Firm initiated
Location
Stockholm, N/A, Sweden

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0848-2022