Cyberonics VNS Therapy AspireHC Model 105 Generator, 26-0007-4000/3. The firm name on the label is CYBERONICS, INC., Houston, TX. Vagus nerve stimulator.
This recall is an expansion of Z-3019-2017 and Z-3020-2017, which was initiated to fix the devices premature battery depletion, caused by electrical leakage on the circuit board assemblies of the Models 105 and 106 generators. *Note this recall occurred in November 2018.
Distribution US nationwide, including Puerto Rico. There was no foreign/military/government consignees.
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0868-2020