FDA recall Z-1105-2020

Intuitive Surgical, Inc. · Class II · device

Product

Da Vinci Xi/X EndoWrist Monopolar Curved scissors (MCS), model number 470179-19, UDI: (00) 886874112298

Reason for recall

This recall is being initiated because lots of EndoWrist Monopolar Curved Scissor instruments are not recognized by the integrated electrosurgical unit, which prohibits the activation of monopolar energy from the surgeon console.

Distribution

US National Distribution Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Georgia, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, Tennessee, Texas, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming OUS:None

Key facts

Status
Terminated
Initiation date
2018-12-04
Report date
2020-02-12
Termination date
2022-08-24
Voluntary/Mandated
Voluntary: Firm initiated
Location
Sunnyvale, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1105-2020