FDA recall Z-1695-2023

Mechatronics USA · Class III · device

Product

RR Mechatronics Control Level A ESR Control4.6 ml QRR049002- In-vitro diagnostic quality control material to monitor the precision of Starrsed Erythrocyte Sedimentation Rate (ESR) instruments and procedures. Product Code: A0026578

Reason for recall

QRR049001 Lot QCAB6AN505 incorrectly labelled as QRR049002 Lot QCAB6AAA28. The incorrectly labelled vial will upload the range for the Starrsed Control Level A and the instrument result will be flagged as out of range.

Distribution

Worldwide distribution - US Nationwide distribution in the state of OH and the country of Netherlands.

Key facts

Status
Ongoing
Initiation date
2023-04-14
Report date
2023-06-14
Voluntary/Mandated
Voluntary: Firm initiated
Location
Warwick, RI, United States

Primary source