Philips Respironics E30 with Humidifier, Continuous Ventilator, Minimum Ventilatory Support, Facility Use; Product Numbers DSX9999H11, INX9999H19, RINX9999H19 The Philips Respironics E30 ventilator is intended to provide invasive and non-invasive ventilatory support for individuals with Respiratory Insufficiency. It is specifically for the care of adult and pediatric patients >7 years of age and >18kgs.
The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the device s the air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.
Global distribution
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1972-2021