FDA recall Z-2167-2026

Hologic, Inc · Class II · device

Product

SELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS. Model Numbers: SDA-SYS-3000-2D-HTC, SDA-SYS-3000-3D, SDM-SYS-6000-2D, SDM-SYS-6000-3D, SDM-00001-2D, SDM-00001-3D, SDM-00001-M2D, SDM-00001-M3D, SDM-05000-2A2, SDM-05000-2A3, SDM-05000-2AC, SDM-05000-2D2, SDM-05000-2D3, SDM-05000-2DC, SDM-05000-3D2, SDM-05000-3D3, SDM-05000-3DC, SDM-SYS-9000-2D, SDM-SYS-9000-3D.

Reason for recall

Complaints have been received of systems developing loose, missing, or broken internal bolts over time.

Distribution

US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2026-04-08
Report date
2026-05-20
Voluntary/Mandated
Voluntary: Firm initiated
Location
Marlborough, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2167-2026