# FDA recall Z-2200-2026

> **MICROVENTION INC.** · Class II · device recall initiated 2026-04-06.

## Product

LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-CAS/4.5 mm x 18 mm x 14 mm, 213517-CAS/3.5 mm x 17 mm x 13 mm

## Reason for recall

Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress, neointimal hyperplasia, vessel injury/rupture; if undersized may lead to incomplete expansion/malposition, acute stent thrombosis, distal embolization, ischemic neurological events

## Distribution

International distribution to the country of China.

## Key facts

- **Recall number:** Z-2200-2026
- **Recalling firm:** MICROVENTION INC.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2026-04-06
- **Report date:** 2026-05-20

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Aliso Viejo, CA, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2200-2026

## Citation

> MORVS. FDA recall Z-2200-2026. Retrieved 2026-11-16 from https://api.morvs.ai/recall/Z-2200-2026. Source: US FDA. Licensed CC0.

---

*[Dataset catalog](/datasets/) · [FDA recalls dataset](/datasets/fda-recalls) · CC0 1.0*