FDA recall Z-2224-2026

Fresenius Kabi USA, LLC · Class I · device

Product

Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004

Reason for recall

Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.

Distribution

Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;

Key facts

Status
Ongoing
Initiation date
2026-05-06
Report date
2026-06-10
Voluntary/Mandated
Voluntary: Firm initiated
Location
North Andover, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2224-2026