FDA recall Z-2245-2026

GE Medical Systems Information Technologies Inc · Class II · device

Product

GE HealthCare CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA, REF #5697603, Product Code KSA; used in conjunction wtih ApexPro CARESCAPE Telemetry Server (CTS) v6.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Reason for recall

GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).

Distribution

Worldwide distribution.

Key facts

Status
Ongoing
Initiation date
2026-04-24
Report date
2026-06-03
Voluntary/Mandated
Voluntary: Firm initiated
Location
Waukesha, WI, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2245-2026