FDA recall Z-2258-2020

GE Healthcare, LLC · Class II · device

Product

System, Tomography Computed Emmission

Reason for recall

GE Healthcare has identified that a small number of Nuclear Medicine systems are operating without currently updated software. In a recent event, a patient s finger was injured during the Unload process while the automatic Unload feature was being used; the software on the system was not current.

Distribution

Domestic Distribution: AL, AR, AZ, CA, CO, CT, DC, DE, FL,GA, IA, IL, IN, KY, LA,MA, MD, ME, MI, MN, MO, MS, NC, NE, NM, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT,VA, WI and WV. Foreign Distribution: Argentina, Australia, Belgium, Brazil, Canada, Colombia, Denmark, France, Germany, Ghana, India, Israel, Italy, Japan, Korea, Malaisya, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Singapore, Sweden, Taiwan, Thailand, Turkey and United Kingdom.

Key facts

Status
Ongoing
Initiation date
2018-11-30
Report date
2020-06-10
Voluntary/Mandated
Voluntary: Firm initiated
Location
Waukesha, WI, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2258-2020