Philips Azurion 7B20/15, Model Numbers: 722068, 722226 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
US Nationwide distribution.
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2289-2026