FDA recall Z-2321-2026

Fresenius Medical Care Holdings, Inc. · Class II · device

Product

Bicarby" Dialysate; Model Numbers: RFP-400-G, RFP-401-G, RFP-402-G, RFP-404-G, RFP-407-G, and RFP-456-G;

Reason for recall

Given the increased number of leaks reported by users, it was determined to be the appropriate course of action to notify users of the potential for the occurrence of Bicarby luer-lock leaks during manipulation of the frangible and tubing when preparing the product for use for treatment. This issue presents the potential of a slip and fall hazard associated with fluid on the floor.

Distribution

US Nationwide distribution in the states of NY, TN and TX.

Key facts

Status
Ongoing
Initiation date
2026-05-12
Report date
2026-06-17
Voluntary/Mandated
Voluntary: Firm initiated
Location
Waltham, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2321-2026