FDA recall Z-2359-2026

Datex-Ohmeda Inc · Class I · device

Product

Panda Freestanding Warmer, Infant Radiant Warmer Product Identifiers: 2063568-001 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.

Reason for recall

Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.

Distribution

Pending

Key facts

Status
Ongoing
Initiation date
2026-05-08
Report date
2026-06-24
Voluntary/Mandated
Voluntary: Firm initiated
Location
Waukesha, WI, United States