# FDA recall Z-2463-2026

> **Hamilton Medical AG** · Class I · device recall initiated 2026-05-29.

## Product

Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.

## Reason for recall

Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement

## Distribution

US Nationwide distribution including in the states of KS, WA, VT, TX, WI, AR, OH, PA, CA, CO, ID, AZ, ME, MO, MI, IL, OR, VA, FL, NJ, SD, NY, NM, TN, CT, NE, OK, AL, GA, KY, NH, WY, ND, IN, UT, WV, DE, IA, PR, MN, LA, AK, HI.

## Key facts

- **Recall number:** Z-2463-2026
- **Recalling firm:** Hamilton Medical AG
- **Classification:** Class I
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2026-05-29
- **Report date:** 2026-07-08

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Domat/Ems, N/A, Switzerland

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2463-2026