# FDA recall Z-2521-2026

> **Intuitive Surgical, Inc.** · Class II · device recall initiated 2026-04-22.

## Product

Da Vinci ASSY, PSS, SP1098, Part Number: 380601

## Reason for recall

Operative components of surgical systems potentially distributed with E-brake retainer pin, so in event of a structural column failure, the E-brake pin would prevent E-brake engagement and may result in uncontrolled motion. In addition, if it moves/backs out it may cause restricted column range of motion and/or uncontrolled motion, which may cause patient tissue injury/crushing injury/bleeding.

## Distribution

Worldwide - US Nationwide distribution including in the states of CA, GA, OH, PA, NY, MN, TN, OR, IL, MD, NJ, AL, NV, TX, FL, LA, CT, NC, AZ, IA, MO, WA, MI, WV, WI, KY, KS, VA, OK, MA, IN, CO, SC, SD, MS, NH, UT, ME and the countries of Belgium, Denmark, Sweden, France, Japan, Germany, Israel, Switzerland, South Korea, Czech Republic, Austria, United Kingdom, Italy, Spain, Ireland, Saudi Arabia, Greece, Taiwan, Australia, China, Turkey, Poland, United Arab Emirates, Hong Kong, Romania.

## Key facts

- **Recall number:** Z-2521-2026
- **Recalling firm:** Intuitive Surgical, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2026-04-22
- **Report date:** 2026-07-01

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Sunnyvale, CA, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2521-2026

## Citation

> MORVS. FDA recall Z-2521-2026. Retrieved 2026-10-24 from https://api.morvs.ai/recall/Z-2521-2026. Source: US FDA. Licensed CC0.

---