FDA recall Z-2548-2026

Baxter Healthcare Corporation · Class I · device

Product

Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing device

Reason for recall

There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

Key facts

Status
Ongoing
Initiation date
2026-05-21
Report date
2026-07-08
Voluntary/Mandated
Voluntary: Firm initiated
Location
Deerfield, IL, United States