# FDA recall Z-2579-2026

> **Philips North America Llc** · Class II · device recall initiated 2026-06-05.

## Product

Philips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions: TELE Monitor 5500, 1.4 Ghz, ECG only, ex; Telemetry Monitor 5500 1.4 GHz FAST EX.

## Reason for recall

Potential for device reset to default "NEW_DEVICE" state, which may lead to a to loss of device configuration and the equipment label, resulting in a failure to reconnect to the PIC iX central monitoring system.

## Distribution

US Nationwide distribution in the states of AZ, CA, CO, IN, MA, MI, MN, PA, SD, TN.

## Key facts

- **Recall number:** Z-2579-2026
- **Recalling firm:** Philips North America Llc
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2026-06-05
- **Report date:** 2026-07-01

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Cambridge, MA, United States