FDA recall Z-2640-2026

Accuray Incorporated · Class II · device

Product

Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator

Reason for recall

In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.

Distribution

Worldwide distribution - US Nationwide and the countries of China, Colombia, Czech Republic, France, Hong Kong, Japan, Korea, Republic of, Latvia, Mexico, Myanmar, Saudi Arabia, Taiwan, Turkey, Ukraine.

Key facts

Status
Ongoing
Initiation date
2026-06-16
Report date
2026-07-08
Voluntary/Mandated
Voluntary: Firm initiated
Location
Madison, WI, United States