FDA recall Z-2641-2026

Siemens Medical Solutions USA, Inc · Class II · device

Product

ARTIS icono biplane. Model Number: 11327600.

Reason for recall

Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.

Distribution

US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2026-05-20
Report date
2026-07-08
Voluntary/Mandated
Voluntary: Firm initiated
Location
Malvern, PA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2641-2026