# FDA recall Z-2644-2026

> **Accuray Incorporated** · Class II · device recall initiated 2026-06-16.

## Product

Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator

## Reason for recall

In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.

## Distribution

Worldwide - US Nationwide and and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Egypt, France, Germany, Greece, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kenya, Korea, Republic of, Liechtenstein, Lithuania, Luxembourg, Macao, Mexico, Netherlands, New Zealand, Pakistan, Poland, Portugal, Qatar, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.

## Key facts

- **Recall number:** Z-2644-2026
- **Recalling firm:** Accuray Incorporated
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2026-06-16
- **Report date:** 2026-07-08

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Madison, WI, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2644-2026