# FDA recall Z-2658-2026

> **Bard Access Systems, Inc.** · Class II · device recall initiated 2026-05-21.

## Product

CK000662	PowerPICC SOLO HF Catheter 5 Fr Triple-Lumen	  UDI-DI Code: 00801741129674      Catheter stylets provide internal reinforcement to aid in catheter placement.

## Reason for recall

Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.

## Distribution

Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.

## Key facts

- **Recall number:** Z-2658-2026
- **Recalling firm:** Bard Access Systems, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2026-05-21
- **Report date:** 2026-07-08

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Salt Lake City, UT, United States