FDA recall Z-2669-2026

Laborie Medical Technologies Corp · Class II · device

Product

Solar GI System (Water Perfusion); Product Name (Catalog Number): Solar GI LGI H2O, 12p, Trolley system (G3-6), Solar GI HRM H2O, 24p, Compact Pole system (G3-7), Solar GI HRM H2O, 24p, Trolley system (G3-8), Solar GI HRM H2O, 36p, Trolley system (G3-12), Solar GI HRAM H2O, 24p, Trolley system (G3-14);

Reason for recall

Under extended clinical use conditions (e.g., long duration studies), moisture generated within the pneumatic circuit may accumulate inside the system's internal tubing or at the hydrophobic filter and waterproof pressure sensors. Accumulated moisture may affect airflow or pressure measurement performance if not properly removed.

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Canada, China, Costa Rica, Croatia, Cyprus, Czech Republic. Denmark, Egypt, France, Germany, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Lithuania, Malaysia, Namibia, Nepal, Netherlands. Oman, Osterreich, Pakistan, Poland, Portugal, Puerto Rico, Qatar, Russia, Russian Federation, Saudi Arabia, Serbia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Tunisia, Turkey, UAE, United Kingdom, and Vietnam.

Key facts

Status
Ongoing
Initiation date
2026-05-21
Report date
2026-07-08
Voluntary/Mandated
Voluntary: Firm initiated
Location
Portsmouth, NH, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2669-2026