FDA recall Z-2702-2026

Fresenius Medical Care Holdings, Inc. · Class II · device

Product

5008X Blood Tubing Set, 24 units per case. Includes the below model numbers and descriptions. 1. 5008X HD Pre-Flush Standard Blood Tubing Set; Model Number: 03-5110-6. 2. 5008X HD Pre-Flush With Twister Blood Tubing Set; Model Number: 03-5160-1. 3. 5008X HD/HDF Standard Blood Tubing Set; Model Number: 03-5300-3. 4. 5008X HD/HDF With CLiC Blood Tubing Set; Model Number: 03-5300-3C. 5. 5008X HD/HDF With Twister Blood Tubing Set; Model Number: 03-5350-8.

Reason for recall

Potential for Online Port Errors due to connection issues when connecting the SafeLine of the 5008X CAREset to the substitution fluid port of the 5008X CAREsystem.

Distribution

US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DC, FL, GA, IL, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NM, NY, OH, PA, TX, VA.

Key facts

Status
Ongoing
Initiation date
2026-05-11
Report date
2026-07-15
Voluntary/Mandated
Voluntary: Firm initiated
Location
Waltham, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2702-2026