# FDA recall Z-2718-2026

> **Olympus Corporation of the Americas** · Class II · device recall initiated 2026-05-05.

## Product

Olympus Toomey Evacuator 30cc with Adaptor. Model Number: 188.

## Reason for recall

Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.

## Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Japan, EU, and Latin America.

## Key facts

- **Recall number:** Z-2718-2026
- **Recalling firm:** Olympus Corporation of the Americas
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2026-05-05
- **Report date:** 2026-07-22

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Center Valley, PA, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2718-2026

## Citation

> MORVS. FDA recall Z-2718-2026. Retrieved 2026-10-21 from https://api.morvs.ai/recall/Z-2718-2026. Source: US FDA. Licensed CC0.

---