# FDA recall Z-2724-2026

> **ICU Medical, Inc.** · Class II · device recall initiated 2026-06-10.

## Product

ICU Medical Plum Solo Precision IV Pump, v1.0.3, Item Number 400010401; infusion pump

## Reason for recall

In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate.

## Distribution

Worldwide distribution - US Nationwide and the country of Canada.

## Key facts

- **Recall number:** Z-2724-2026
- **Recalling firm:** ICU Medical, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2026-06-10
- **Report date:** 2026-07-22

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Lake Forest, IL, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2724-2026

## Citation

> MORVS. FDA recall Z-2724-2026. Retrieved 2026-10-17 from https://api.morvs.ai/recall/Z-2724-2026. Source: US FDA. Licensed CC0.

---

*[Dataset catalog](/datasets/) · [FDA recalls dataset](/datasets/fda-recalls) · CC0 1.0*