FDA recall Z-2727-2026

GE Medical Systems, LLC · Class II · device

Product

Optima XR220amx, Mobile Digital-Ready Radiographic System

Reason for recall

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.

Distribution

Worldwide - US Nationwide and the countries of Brazil, Canada, Korea, Mexico, Saudi Arabia, Spain, UK.

Key facts

Status
Ongoing
Initiation date
2026-06-05
Report date
2026-07-22
Voluntary/Mandated
Voluntary: Firm initiated
Location
Waukesha, WI, United States