# FDA recall Z-2957-2026

> **Kico Knee Innovation Company** · Class II · device recall initiated 2026-07-13.

## Product

Brand Name: Kico Knee Innovation  Product Name: ARVIS SURGICAL NAVIGATION SYSTEM GEN 2  Model/Catalog Number: IN-29200  Software Version: ARVIS Hip and Knee Gap Balance Software v3.3.2 (IN-28200)  ARVIS Shoulder Software v2.0.3 (IN-23630)  Product Description: The ARVIS system is a computer-controlled navigation system intended to provide intra-operative measurements to the surgeon to aid in selection and positioning of orthopedic implant components.          The ARVIS Surgical Navigation System is indicated for assisting the surgeon in the positioning and alignment of implants relative to reference alignment axes and landmarks in stereotactic orthopedic surgery. The system aids the surgeon in making intraoperative measurements such as changes in leg length in Hip Arthroplasty. The system is compatible with straight acetabular impactors and with specific offset impactors, identified in the instructions for use, for which an adapter has been validated. Example orthopedic surgical procedures include but are not limited to:      " Total Knee Arthroplasty      " Unicompartmental Knee Arthroplasty: Tibial Transverse Resection      " Hip Arthroplasty          The ARVIS" head mounted display is for displaying augmented reality visualization and information to the user intraoperatively. The augmented/virtual displayed information should not be relied upon solely for absolute positional/alignment information and should always be used in conjunction with the displayed stereotaxic information.  Component: NA

## Reason for recall

Due to an internal sensor issue, image displayed with eyepiece may appear tilted relative to the eyepiece housing

## Distribution

Consignee Consignee Number  Region Device  Quantity RFK MEDICAL  CE102 RICHMOND VA  IN-29200, ARVIS SURGICAL NAVIGATION 1 VALOR MEDICAL  CE118 SAN ANTONIO TX  IN-29200, ARVIS SURGICAL NAVIGATION  2 JOINT VENTURES MEDICAL CE121 OSHKOSH WI  IN-29200, ARVIS SURGICAL NAVIGATION  1 CASTLE MEDICAL  CE122 WATERLOO IA  IN-29200, ARVIS SURGICAL NAVIGATION  6 GREAT LAKES MEDICAL GL108 NAPERVILLE IL   IN-29200, ARVIS SURGICAL NAVIGATION  2 CERTUS MEDICAL  GL112 INDIANAPOLIS IN   IN-29200, ARVIS SURGICAL NAVIGATION  5 I&S MEDICAL  GL113 LOUISVILLE KY   IN-29200, ARVIS SURGICAL NAVIGATION  2 PROCORE MEDICAL MA102 PIEDMONT SC  IN-29200, ARVIS SURGICAL NAVIGATION  1 FOX MEDTECH  MW001 ENGLEWOOD CO   IN-29200, ARVIS SURGICAL NAVIGATION  2 MEDSOURCE  NE131 TIVERTON RI   IN-29200, ARVIS SURGICAL NAVIGATION  4 IMPLANT MEDICAL SALES NE190 HICKSVILLE NY   IN-29200, ARVIS SURGICAL NAVIGATION  2 KINETIC MEDICINE NW105  OMAHA NE   IN-29200, ARVIS SURGICAL NAVIGATION  4 CALSURG NW115 SANTA BARBARA CA  IN-29200, ARVIS SURGICAL NAVIGATION  4 MIDWEST MEDICAL RESOURCES SC181 OVERLAND PARK KS  IN-29200, ARVIS SURGICAL NAVIGATION  3 ORTHO & SURGICAL SOLUTIONS SE175 GUAYNABO PR  IN-29200, ARVIS SURGICAL NAVIGATION  3 JOINT MEDICAL SOLUTIONS SE180 BIRMINGHAM AL  IN-29200, ARVIS SURGICAL NAVIGATION  4 FLT ORTHOPEDICS SE216 TAMPA FL   IN-29200, ARVIS SURGICAL NAVIGATION  6 MAC MEDICAL  W180 GILBERT AZ   IN-29200, ARVIS SURGICAL NAVIGATION  1 CP ORTHO  W214 HONOLULU HI   IN-29200, ARVIS SURGICAL NAVIGATION  1 UNIFIED ORTHOPEDICS W223 RIVERSIDE CA   IN-29200, ARVIS SURGICAL NAVIGATION  18 MOVEMENT MEDICAL W225 SEATTLE WA   IN-29200, ARVIS SURGICAL NAVIGATION  6 Total number of devices in the field 78

## Key facts

- **Recall number:** Z-2957-2026
- **Recalling firm:** Kico Knee Innovation Company
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2026-07-13
- **Report date:** 2026-08-26

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Pymble, Australia