FDA recall Z-2991-2026

Medline Industries, LP · Class II · device

Product

Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product Number/SKU 780-19); breathing circuit

Reason for recall

Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.

Distribution

Worldwide distribution - US Nationwide and the country of Panama.

Key facts

Status
Ongoing
Initiation date
2026-07-13
Report date
2026-09-02
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northfield, IL, United States