FDA recall Z-3105-2026

Cardinal Health 200, LLC · Class II · device

Product

Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, CHONY ORTHO MAJOR SPINE , Catalog Number: POV2SCCPP; 2) KIT,ORTHOPLASTIC MICROVASCULAR, Catalog Number: PWVMFFABLI

Reason for recall

The Urinary Catheter Trays contain CliniClean Pericare Wipes which may potentially be contaminated with a lubricant used during the manufacturing process. If a contaminated wipe is used on a patient, localized skin effects may occur. These may include irritation, erythema, or mild inflammation.

Distribution

US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2026-07-29
Report date
2026-09-09
Voluntary/Mandated
Voluntary: Firm initiated
Location
Waukegan, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3105-2026