FDA recall Z-3110-2026

Lutronic Corporation · Class II · device

Product

Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser energy at wavelengths of 532 nm and 1064 nm through a fiber-optic handpiece. The system may be configured with an Intelligent Cooling Device (ICD) to provide epidermal cooling during laser treatment

Reason for recall

Utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in patient burn.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada and European Union.

Key facts

Status
Ongoing
Initiation date
2026-08-06
Report date
2026-09-16
Voluntary/Mandated
Voluntary: Firm initiated
Location
Goyang, Korea (the Republic of)

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3110-2026