# FDA recall Z-3132-2026

> **Nova Biomedical Corporation** · Class I · device recall initiated 2026-07-29.

## Product

StatStrip Glucose and B- Ketone meter. Model No: 66429. Glucose test system.

## Reason for recall

A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode   including patient identification (ID) barcodes and quality control (QC) solution ID barcodes   when a new barcode is scanned, resulting in the prior barcode information remaining displayed on the screen for the current test. When this defect occurs and the operator accepts the incorrectly displayed patient information without recognizing the mismatch, the current patient's glucose result will be incorrectly assigned to the previous patient's medical record and transmitted to the data management system (DMS), and the current patient's medical record will not receive a result, leaving their glucose status unrecorded and their clinical management without the information needed to guide treatment decisions. The previous patient, whose medical record now contains a glucose result that does not belong to them, may receive inappropriate clinical intervention based on that erroneous result, particularly in critically ill patients, neonates, and infants who cannot communicate symptoms or self-treat. The defect also affects QC barcode scanning: if a QC scan triggers the same retention behavior, QC results may be attributed to an incorrect identifier or QC tests may fail without the user recognizing the underlying cause. This defect does not affect the analytical accuracy of glucose measurements.

## Distribution

Worldwide distribution: US (nationwide) and OUS (foreign) countries of: Austria, Belgium, Bolivia, Brazil, Canada, Cayman Islands, Chile, Croatia, Dominican Rep., Finland, France, Germany, Ireland, Israel, Italy, Netherlands, New Zealand, South Korea, Switzerland, Thailand, Turkey, United Kingdom.

## Key facts

- **Recall number:** Z-3132-2026
- **Recalling firm:** Nova Biomedical Corporation
- **Classification:** Class I
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2026-07-29
- **Report date:** 2026-09-23

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Waltham, MA, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3132-2026

## Citation

> MORVS. FDA recall Z-3132-2026. Retrieved 2026-10-21 from https://api.morvs.ai/recall/Z-3132-2026. Source: US FDA. Licensed CC0.

---