FDA recall Z-3140-2026

CooperSurgical, Inc. · Class I · device

Product

INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP, Replacement Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-0712, 44-0707, 44-0709, 44-0715.

Reason for recall

Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.

Distribution

Nationwide distribution to CT, FL, GA, IA, IL, IN, MD, MI, MS, NJ, NY, OH, PA, VA. International distribution to Brazil and Peru.

Key facts

Status
Ongoing
Initiation date
2026-08-11
Report date
2026-09-30
Voluntary/Mandated
Voluntary: Firm initiated
Location
Trumbull, CT, United States

Primary source