# FDA recall Z-3188-2026

> **Boston Scientific Corporation** · Class II · device recall initiated 2026-07-22.

## Product

Boston Scientific Corporation  StoneSmart Connect Console  Contents: (1) Power Cord, (1) Power Adaptor, (1)DVI Cable, (1) DVI to HDMI Cable, (1) DisplayPort Cable  REF: M0067933000, M006793300H0, M006793300R0, M006793300D0, M006793300Z0, M0067930000, M006793000R0

## Reason for recall

A subset of power adapters used with the connect console may exhibit higher-than-expected resistance in the electrical grounding path. As a result, the power adaptor may fail electrical safety testing during installation.

## Distribution

Worldwide - US Nationwide distribution including in the states of AZ, FL, CA, OR, IL, NC, NE, UT, CO, OH, AR, NM, AL, WI, TN, IN, GA, OH, WI, KY, CT, KS, SC, IL, NE, MO, MS, DE, TX, NY, WV, MA, MT, WA, WY, NV, HI, ID, MO, ME, WV, NE, OK, MN, LA, WV, NV, PA, NH, NJ, MI, MO, AK, VA, MS, WY, ME and the countries of Australia, Finland, France, Germany, Great Britain, Greece, Ireland, Italy, Lebanon, Norway, Poland, Portugal, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Turkiye, Canada, Hong Kong, Singapore, Taiwan, Argentina, Chile, Japan.

## Key facts

- **Recall number:** Z-3188-2026
- **Recalling firm:** Boston Scientific Corporation
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2026-07-22
- **Report date:** 2026-09-23

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Marlborough, MA, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3188-2026

## Citation

> MORVS. FDA recall Z-3188-2026. Retrieved 2026-10-18 from https://api.morvs.ai/recall/Z-3188-2026. Source: US FDA. Licensed CC0.

---

*[Dataset catalog](/datasets/) · [FDA recalls dataset](/datasets/fda-recalls) · CC0 1.0*