# Electrocution, electric shock at Boeing (LMTF)
> OSHA SIR `2025066082` · 6/25/2025 · OGDEN, UTAH
## Severity
- **1 hospitalization**
- **Body part:** BODY SYSTEMS
- **Nature of injury:** Electrocution, electric shock
## Cause
- **Event:** Direct exposure to electricity  unspecified
- **Source:** Medical imaging and radiological machinery, equipment
- **Secondary source:** Secondary source not applicable
## Employer
- **Name:** Boeing (LMTF)
- **Location:** 12000 W 900 S St, OGDEN, UTAH, 84404
- **NAICS:** 541380
- **Jurisdiction:** 0
## Narrative

An X-ray machine had lost power and an employee went to replace the power supply. While replacing a cable on the power supply assembly, the employee's right hand contacted the grounding terminals coming out of the back of the power supply assembly and they received an electric shock. The employee was hospitalized.

---