FDA recall Z-2200-2026

MICROVENTION INC. · Class II · device

Product

LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-CAS/4.5 mm x 18 mm x 14 mm, 213517-CAS/3.5 mm x 17 mm x 13 mm

Reason for recall

Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress, neointimal hyperplasia, vessel injury/rupture; if undersized may lead to incomplete expansion/malposition, acute stent thrombosis, distal embolization, ischemic neurological events

Distribution

International distribution to the country of China.

Key facts

Status
Ongoing
Initiation date
2026-04-06
Report date
2026-05-20
Voluntary/Mandated
Voluntary: Firm initiated